XZ-P5 vacuum blood collection tube special capping centrifuge manual

First, the model name

XZ-P5 low speed centrifuge ( vacuum blood collection tube special decapping centrifuge )

Second, use

XZ-P5 low-speed centrifuge vacuum blood collection tube special capping centrifuge can be widely used in clinical medicine, biochemistry, immunology and other fields is a routine instrument for centrifugal sedimentation in the laboratory.

Third, technical parameters

Maximum speed: 4000r/min
Maximum centrifugal acceleration: 3050×g
Timing range: 0~99min
Speed ​​accuracy: ±30r/ rpm
Noise: 60dB
Appearance size: 440 × 550 × 330mm
Weight: 30kg
Can be equipped with a rotor:

Rotor number

Maximum speed

Maximum capacity

Maximum centrifugal force

No.1 horizontal rotor

4000r/min

48x7ml

3050xg

No.2 horizontal rotor

4000r/min

48x5ml

3050xg

Fourth, the principle and characteristics

The XZ-P5 capping centrifuge is driven by a motor directly, and the special damper is used for vibration reduction. The rotor is precision casted with high-strength materials, and the rotor is easy to handle. Microcomputer control, AC variable frequency motor, high precision, low noise. Large screen LCD display, speed, centrifugal force, running time, fault display. The unique design of the double layer of the adapter prevents the cap from falling out during the operation of the centrifuge, and automatically removes the cap to improve the efficiency of the experiment. The rise/deceleration time is short (30S) with 1 to 8 liters/downshift. With electronic door lock, overspeed, automatic rotor identification, unbalanced detection and protection to ensure personal safety. Touch panel, programmable operation, automatic conversion of rotational speed centrifugal force, easy to operate. It is suitable for centrifugation in hospital biochemistry, radiology laboratory, and central blood station biological products.

Fifth, the installation

Installer:
(1) The table top is solid and flat, preventing the capping centrifuge from vibrating or moving the instrument during operation.
(2) The power supply has an independent ground line. It is strictly forbidden to share the neutral line and the ground line to prevent electric shock and injury.
(3) There should be no corrosive gas and electromagnetic field interference around the capping centrifuge to prevent premature failure of the external protective layer of the instrument and interference with electrical components.
(4) Place the capping centrifuge to ensure that the four rubber feet are evenly stressed.

Six, the control panel introduction

Cap removal centrifuge 1, select the settings you need
2, increase the button
3, function parameter drop button
4, confirm the function parameter approval key
5, start the centrifuge speed button
6,
centrifuge
speed stop button
7, open the door cover button
8, speed and centrifugal force display switch button
9, display failure display:
E1=Unbalance E2=Overspeed E3=Door switch (start)
E5=Time setting is zero E6=Temperature
E7=Motor does not turn E8=Communication E9=System

Seven, the operation process

Operating procedure
1. Turn on the power, press the door lock button to open the door automatically, and the centrifugal tube in the centrifugal sample loading area is placed symmetrically in the rotor, and each tube sample can be visually observed.
2. Press the SET button to flash the rotor number. The first one is the rotor number. Press the +- button, then press the SET button. Press the +- button for the pre-selected speed.
3. Parameter setting Press the setting button, the rotor number window will flash, press the plus or minus button to select the rotor number;
Press the set button speed/centrifugal force window to flash and press the plus or minus button to select the speed;
Press the Setup button to raise/lower the window and press the plus or minus button to select the up/down.
Press the setting button time window to flash and press the plus or minus button to select the time;
The selected parameters are correct, press the confirmation button, which is the parameter you are currently setting.
4. When your parameters are selected, press the approval button again. During the operation, you can directly display the centrifugal force and press the speed/centrifugal force switch button. (Note: the speed can still be modified during the operation.) Press the start button to start working. .
Time to zero or press the stop button, after the speed is zero, press the door lock button to automatically open the door.

Anqiu Ginger is famous for its bright yellow skin, good fruity shape, compact structure and strong ginger flavor. In addition to containing a large number of carbohydrates, proteins, vitamins and minerals, it also contains a large number of gingerol, gingerolone, gingerol and gingerol, etc., which has high edible value and medicinal value.