XZ-P4 vacuum blood collection tube special capping centrifuge manual

First, the product model, name


XZ-P4 low speed centrifuge (vacuum blood collection tube special vacuum blood collection tube capping centrifuge)


Second, use and scope of use


XZ-P4 vacuum blood collection tube special capping centrifuge (hereinafter referred to as the machine) can be widely used in clinical medicine, biochemistry, immunology, blood station and other fields. It is a routine instrument for centrifugal sedimentation in the laboratory.

Third, the main specifications and technical parameters

Maximum speed: 4000r/min
Maximum centrifugal acceleration: 3840×g
Speed ​​accuracy: ±30r/ rpm
Timing range: 0~99min
Machine noise: 65dB;
Weight (without turning head): 150kg
Appearance size: 630 × 730 × 1060 (L × W × H)

Fourth, the rotor configuration

Rotor number

Maximum speed

Maximum capacity

Maximum centrifugal force

1 horizontal rotor

4000r/min

96x7ml/5ml

3840xg

2 horizontal rotor

4000r/min

148x7ml/5ml

3840xg

V. Working principle and characteristics

The machine is directly driven by a variable frequency motor, and the special damper is used for vibration reduction. The rotor is formed by high-strength steel, and the rotor is easy to handle. The floor-standing centrifuge has a hood and low noise. The amount of sample processed is large. Microcomputer control, AC variable frequency motor, precise speed, high screen LCD display, speed, centrifugal force, running time, fault display. The unique design of the double layer of the adapter prevents the cap from falling out during the operation of the centrifuge, and automatically removes the cap to improve the efficiency of the experiment. The rise/deceleration time is short (30S) with 1 to 8 liters/downshift. With electronic door lock, overspeed, automatic rotor identification, unbalanced detection and protection to ensure personal safety. Touch panel, programmable operation, automatic conversion of rotational speed centrifugal force, easy to operate. It is suitable for centrifugation in hospital biochemistry, radiology laboratory, and central blood station biological products. This machine can replace the rotor of a variety of pipe racks to meet the needs of different users.

Sixth, installation

Installer:
(1) The ground is firm and flat to prevent the instrument from vibrating or moving the instrument during operation.
(2) Adjust the left and right foot screws to ensure the level of the whole machine.
(3) With independent ground wire, it is strictly forbidden to share the zero wire and the ground wire to prevent electric shock and injury.
(4) There should be no corrosive gas and electromagnetic field interference around the instrument to prevent premature failure of the external protective layer of the instrument and interference with electrical components.

Seven, operation


Cap cap centrifuge control panel
control panel

Cap removal centrifuge

1. Rotor display, select the rotor number you need.
2. The speed display speed display window is 5 digits. The display window can display two functions. The first function: display the speed parameter, and the second function: display the centrifugal force parameter.
3. Temperature display The temperature display window is a three-digit number. The window displays the temperature parameter. The temperature control accuracy is ±2°C with 0.1 degree step.
4. The time/fault display time display window is three digits, with 1 second for 1 minute, the longest time for 99 minutes, and the centrifugal process time decreasing display.
Fault display:
E1=Unbalance E2=Overspeed E3=Door switch (start)
E4=door switch (in operation) E5=blocking
E6=Setting error E7=The plug loosening function display setting, speed, temperature, time and call, centrifugal force and fault display.
5. Operate the button function Press the start button to start the centrifuge speed, the stop button is the centrifuge speed stop, press the stop button to open the machine after the machine stops, select the setting button for the start of the function setting. Add the key to the function parameter up button.
The down key is the function parameter down button, and the confirmation key is the function parameter approval button.

Eight, operation and use

1. Open the door cover and press the stop button to open the centrifuge cover automatically.
2. Install the rotor (1) Before each use, carefully check whether the rotor is cracked or not, and whether there are corrosive points (especially the bottom of each centrifugal hole of the angle rotor, the bottom of the cup hole of the horizontal rotor). It is strictly forbidden to use a hanging cup or an angular rotor with cracked spots. It is strictly forbidden to use a rotor that exceeds the shelf life.
(2) Place the rotor body on the rotating shaft of the centrifuge, and let the center shaft of the rotor be screwed into the screw hole of the shaft with the center screw.
3. Turn on the power supply After confirming that the above installation is correct, check whether the power supply is independent of the ground wire, plug in the power supply, turn on the power switch of the main unit, and the digital tube of the display window lights up, and the parameters can be set.
4. Close the door cover and close the door cover. When the latch enters the lock hole, it will make a “beep” sound, and the door cover is covered.
5. Parameter setting Press the selection key to rotate the rotor number window, press the plus or minus key to select the rotor number;
Press the SELECT button to rotate/centrifugal force window and press the plus or minus button to select the speed;
Press the SELECT button to scroll through the temperature window and press the plus or minus button to select the temperature;
Press the SELECT button to flash the time window to select the time;
The selected parameters are correct, press the approval button, which is the parameter you are currently setting.
6, start running:
(1) When the start button is pressed, the speed starts to rise until the speed is preselected and runs at a constant speed. The speed display value is the actual speed of the rotor, and the temperature display value is the actual temperature of the centrifugal chamber.
(2) When it is necessary to track the centrifugal force, press the speed / centrifugal force button to display the centrifugal force.
(3) When the time is displayed as 0 or when the stop is needed during the running, the motor signal control line is cut off and the speed starts to decrease. When the speed is displayed as 0, the buzzer prompts to open the centrifuge chamber cover automatically. Then take out the centrifugal pattern. If the next running program (parameters such as temperature, time, speed, etc.) does not change, select the programmed button to start.
(4) If there is a malfunction during operation, the centrifuge will automatically stop.
(5) When using a PE centrifuge tube, the centrifugation temperature should not be lower than 4 °C to prevent the centrifuge tube from brittle cracking.
(6) When the centrifuge suddenly loses power during operation, you need to take out the sample immediately. There is a small hole on the left side of the machine. Gently insert the emergency lever (T-plate wrench) into the clockwise rotation of the centrifuge door to open and remove. sample. As shown below
Cap removal centrifuge

Nine, warranty terms

The company solemnly promises that from the date the user purchases the product from our company (hereinafter referred to as the manufacturer), the user enjoys the following product after-sales warranty service:
1. This product is subject to a 12-month free warranty period from the date the user purchases from the manufacturer;
2. This product is issued within one month from the date the user purchases from the manufacturer and the quality problem, the manufacturer refunds, replaces, repairs;
3. The product has been replaced and repaired by the manufacturer within three months from the date of purchase by the user from the manufacturer;
4. The product enjoys a paid lifetime service from the date the user purchases from the manufacturer;
5, the product failure caused by the following reasons is not the manufacturer's 12-month free warranty service:
(1) The user does not operate correctly according to the instruction manual;
(2) The user has not communicated with the manufacturer to repair it by himself, and the product is modified to cause product failure;
(3) The user fails to use the technical parameter range to cause product failure;
(4) The user malfunctions due to poor use of the environment;
(5) Product damage caused by force majeure such as earthquakes, lightning strikes, fires, feng shui disasters, abnormal voltages or natural disasters;
(6) The product is damaged due to improper selection of the transportation method after the user purchases the product.
Related Products: xz-p4 capping centrifuge


In a Hydroponic System, plants are placed in nutrient-enriched water. Mediums such as gravels, sand or vermiculite can also be used. Thus plants which are placed in the water absorb the nutrients very easly. When the nutrients from the water starts finishing , water is recycled or additional nutrients are added.